1.4Which characteristics, for which test
What is happening
The procedure's purpose decides which performance characteristics are relevant. ICH Q2 tabulates this; USP ⟨1225⟩ groups procedures into four categories.
Worth memorising
Identification: specificity. Limit test: specificity + DL. Quantitative impurity: specificity, range with QL, accuracy, precision. Assay: specificity, range, accuracy, precision.
At the bench
Open the planner tool, pick your procedure type, and use its output as the skeleton of the protocol's characteristic table.
The ICH Q2 matrix in brief. ✓ = normally evaluated; — = not normally evaluated; (✓) = depends on the case.
| Characteristic | Identification | Impurity — quantitative | Impurity — limit test | Assay / content / potency |
|---|---|---|---|---|
| Specificity / selectivity | ✓ | ✓ | ✓ | ✓ |
| Response (linearity) | — | ✓ | — | ✓ |
| Range | — | ✓ | — | ✓ |
| Accuracy | — | ✓ | — | ✓ |
| Repeatability | — | ✓ | — | ✓ |
| Intermediate precision | — | ✓ | — | ✓ |
| Detection limit (DL) | — | (✓) | ✓ | — |
| Quantitation limit (QL) | — | ✓ | — | — |
Q2(R2) groups response and the lower range limit (DL and QL) under “range”, reflecting that they describe where in the concentration scale the procedure is usable. The older Q2(R1) listed linearity, range, DL and QL separately. The science is unchanged; the organisation is clearer. Robustness is still expected, but Q2(R2) and Q14 place its study in development.
USP ⟨1225⟩ categories
| Category | Procedures | Main characteristics |
|---|---|---|
| I | Assay of major components, including preservatives | Accuracy, precision, specificity, linearity, range |
| II | Impurities and degradants — quantitative or limit test | Quantitative: all except DL; limit test: specificity and DL |
| III | Performance characteristics (e.g. dissolution, drug release) | Precision; others as the procedure requires |
| IV | Identification | Specificity |
Validation planner
Choose the procedure type to generate a characteristic-by-characteristic plan with a minimum design and a starting acceptance criterion for each.
| Characteristic | Status | Minimum design | Starting criterion — edit to your specification |
|---|---|---|---|
| Specificity | Required | Blank, placebo, individual impurities, spiked placebo, 5 stress conditions, purity + orthogonal | |
| Response (linearity) | Required | ≥ 5 levels, QL to ≥ 120 % of spec, independent stocks | |
| Range | Required | QL to 120 % of spec | |
| Accuracy | Required | QL, 100 %, 120 % of spec × 3 preparations | |
| Repeatability | Required | 6 preparations at the spec level (or 3 × 3) | |
| Intermediate precision | Required | ≥ 6 runs, analysts, days, instruments; ANOVA | |
| QL | Required | Estimate (10σ/S or S/N 10), confirm with 6 preparations | |
| DL | Conditional Useful for peak-detection decisions | 3.3σ/S or S/N 3 | |
| RRF | Conditional Where impurities are quantified against the API | Slopes of impurity vs API in the same run | |
| Robustness | Required | From development; critical-pair resolution as response | |
| Solution stability | Required | Degradant growth over hold time |
Starting criteria are common industry conventions, not ICH requirements — every one must be justified against your specification and intended use. Ctrl+P prints this table cleanly as a protocol skeleton.