The ATP and the reportable range
Part 1 · Foundations & the Lifecycle · 10 min
Part 1 · Foundations & the Lifecycle

1.3The ATP and the reportable range

Write the target before you aim. The ATP is the method's job description — and it says nothing about HPLC.

What is happening

The ATP fixes WHAT must be measured and HOW WELL, independent of technique. The reportable range fixes WHERE across the concentration scale that performance must hold.

Worth memorising

Classic Q2 minimum ranges: assay 80–120 %, content uniformity 70–130 %, impurities from the reporting threshold to 120 % of the specification.

At the bench

For every quantitative method, draw the range as a bar: specification limits in the middle, the reportable range outside them, the QL at the bottom for impurities.

A usable ATP reads like a specification for the measurement itself: “The procedure must quantify impurity B in the drug product from 0.05 % to 0.30 % w/w relative to the label claim, such that a reportable result falls within ±15 % of the true value with at least 95 % probability.” That sentence already tells you the reportable range, the accuracy and precision you need, and — if you think about it — the QL the method must achieve.

Assay80 %120 %spec 95–105 %Content uniformity70 %130 %spec 85–115 %Impurity Blimit 0.20 %RT 0.05 %120 % = 0.24 %QLSolid bar: specification. Shaded band: minimum reportable range. RT = reporting threshold. The QL needs margin below it.
Reportable ranges drawn against specifications. The range must extend beyond the limits, because results near a limit are exactly the ones that matter.

Minimum reportable ranges

Minimum ranges long recommended in ICH Q2 for common tests. Confirm the exact wording against the current guideline before writing them into a protocol.

Use of the procedureMinimum reportable rangeWhy
Assay of drug substance or product80 % to 120 % of the declared content or the specificationResults near either limit must be reliable
Content uniformity70 % to 130 % of the declared contentIndividual units vary more than batches
DissolutionBeyond the specified range at both ends (the long-standing Q2 convention is ±20 %)Early timepoints can be low; late ones can exceed label
Impurity (quantitative)Reporting threshold to 120 % of the specificationThe method must measure from the first reportable value to past the limit
Assay and impurities in one runReporting threshold of the impurity to 120 % of the assay specificationOne calibration has to cover both jobs

Where acceptance criteria come from

Acceptance criteria are not copied from the last protocol. They follow from the ATP and from the specification the method will be used against. A useful check: if the method's total error were as large as your acceptance criteria allow, would a batch at the specification limit still be classified correctly most of the time? If not, the criteria are too loose for the job, however conventional they look.