1.3The ATP and the reportable range
What is happening
The ATP fixes WHAT must be measured and HOW WELL, independent of technique. The reportable range fixes WHERE across the concentration scale that performance must hold.
Worth memorising
Classic Q2 minimum ranges: assay 80–120 %, content uniformity 70–130 %, impurities from the reporting threshold to 120 % of the specification.
At the bench
For every quantitative method, draw the range as a bar: specification limits in the middle, the reportable range outside them, the QL at the bottom for impurities.
A usable ATP reads like a specification for the measurement itself: “The procedure must quantify impurity B in the drug product from 0.05 % to 0.30 % w/w relative to the label claim, such that a reportable result falls within ±15 % of the true value with at least 95 % probability.” That sentence already tells you the reportable range, the accuracy and precision you need, and — if you think about it — the QL the method must achieve.
Minimum reportable ranges
Minimum ranges long recommended in ICH Q2 for common tests. Confirm the exact wording against the current guideline before writing them into a protocol.
| Use of the procedure | Minimum reportable range | Why |
|---|---|---|
| Assay of drug substance or product | 80 % to 120 % of the declared content or the specification | Results near either limit must be reliable |
| Content uniformity | 70 % to 130 % of the declared content | Individual units vary more than batches |
| Dissolution | Beyond the specified range at both ends (the long-standing Q2 convention is ±20 %) | Early timepoints can be low; late ones can exceed label |
| Impurity (quantitative) | Reporting threshold to 120 % of the specification | The method must measure from the first reportable value to past the limit |
| Assay and impurities in one run | Reporting threshold of the impurity to 120 % of the assay specification | One calibration has to cover both jobs |
Where acceptance criteria come from
Acceptance criteria are not copied from the last protocol. They follow from the ATP and from the specification the method will be used against. A useful check: if the method's total error were as large as your acceptance criteria allow, would a batch at the specification limit still be classified correctly most of the time? If not, the criteria are too loose for the job, however conventional they look.