1.1What validation actually proves
What is happening
You are collecting planned, pre-judged evidence that the reportable result is close enough to the truth, often enough, for the decision that result will drive.
Worth memorising
Fitness for purpose = the right characteristics, over the right range, against acceptance criteria written BEFORE the data exist.
At the bench
Before touching the instrument, write the one sentence the whole study serves: what is measured, in what, over what range, to support which decision.
A validation study is an experiment with a verdict decided in advance. You state what the method must achieve, you design the experiments that would reveal a shortfall, you run them, and you compare what you found with what you said you needed. The comparison is the validation. Without acceptance criteria set beforehand, the same data are just a characterisation — interesting, but not evidence of anything.
This is why the order of work matters so much. A laboratory that runs the experiments and then writes criteria that the data happen to meet has not validated anything; it has described its method in flattering terms. Auditors know this pattern well, and the first thing many of them check is whether the protocol was approved before the first validation injection.
What validation is not
Five activities that get confused with one another, and the question each one answers.
| Activity | The question it answers | Where it sits |
|---|---|---|
| Development | What conditions give a separation that works, and how far can they move? | Before validation — ICH Q14 |
| Validation | Does the finished procedure meet pre-defined criteria for its purpose? | ICH Q2(R2), USP ⟨1225⟩ |
| Verification | Does an already-validated (e.g. compendial) procedure work in my lab, on my sample? | USP ⟨1226⟩ |
| Transfer | Can a receiving lab produce results equivalent to the sending lab? | USP ⟨1224⟩ |
| System suitability | Is the system working properly today, before I trust this run? | Every run — USP ⟨621⟩, Ph. Eur. 2.2.46 |
The reportable result is the thing being validated
What reaches the certificate of analysis is not a peak area. It is a reportable result: perhaps the mean of two preparations, each injected twice, calculated against a standard. Validation should mimic that replication scheme, because the precision of one injection and the precision of the reportable result are different numbers. ICH Q2(R2) makes this explicit: the characteristics are evaluated for the procedure as it will be used.